Ad-weighing ML-50 Manuale Utente Pagina 17

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 75
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 16
MS-70 / MX-50 / MF-50 / ML-50 15
6.2.
6.2.6.2.
6.2. Setting the Clock and C
Setting the Clock and CSetting the Clock and C
Setting the Clock and Calendar
alendaralendar
alendar
Adjust the built-in clock and calendar before use.
6.2.1.
6.2.1.6.2.1.
6.2.1. Operation
OperationOperation
Operation
1. Turn on the analyzer.
The gram unit (of weighing mode) is displayed.
2. Press and hold the SELECT key to display Cl adj .
3. Press the ENTER key to display the calendar.
Example: 15th April, 2002
4. To skip the calendar settings.
Press the or key to proceed step 5.
To adjust the calendar settings.
Press SELECT key. Adjust the calendar using the
following keys.
SELECT key..... Selects a figure.
, key .............. Selects a value for the figure.
ENTER key ...... Stores the current date and
proceeds to step 5.
RESET key....... Cancels the adjustment and
proceeds to step 5.
Symbols and arrangement of the calendar
ymd ................... Year, month, day
mdy ................... Month, day, year
dmy ................... Day, month, year
The arrangement of the calendar is used for the report
of GLP, GMP and ISO.
5. Time is displayed.
6. To finish the adjustment.
Press the RESET key to proceed step 7.
To adjust the clock.
Press the SELECT key. Adjust the clock using the
following keys.
SELECT key..... Selects a figure.
, key .............. Selects a value for the figure.
ENTER key ...... Stores time and proceeds to
step 7.
RESET key....... Cancels the adjustment and
proceeds to step 7.
7. When finishing the adjustment, dp is displayed.
Press the RESET key to return to the weighing mode.
0.000g
Cl adj
ymd
02.04.15
12:34:56
ymd
02.04.15
dp
0.000g
12:34:56
To time
Date adjustment
Time adjustment
Vedere la pagina 16
1 2 ... 12 13 14 15 16 17 18 19 20 21 22 ... 74 75

Commenti su questo manuale

Nessun commento